“史賽克”新格導引線安全警訊
| 發布日期:2024-05-08 | 更新日期:2024-05-08 發布單位:醫療器材及化粧品組

史賽克”新格導引線

安全警訊 

許可證字號:衛署醫器輸字第020895

產品英文名稱:Stryker Synchro Guidewires

受影響規格/型號/批號,及其UDI-DI

型號

批號

UDI-DI

M00326110

0000073409000006756600001167020000174768000009481800000878700000086337000014692800001008630000133181000010989900000979510000130824000017647400000827350000121695000012571500001097820000130864000007760800000883670000137242

04546540688736

M00326010

00001628900000153606000017477000000692760000146147000014573800001461460000175634000017563600001235290000092396000007520700000829650000135804000017227300001457390000078371

04546540688729

M00326310

0000135805000011888600001001820000116675000014787900001285250000067169000014798700000682470000108962

04546540688743

M00326420

00000966290000071378000011559900000704230000066471000006576100001807970000180798

04546540688774

M00326410

0000152226000014896200001088150000138237000012579000000840990000151850000010896100000796940000066472

04546540688767

發布對象:醫療從業人員/醫療器材專業人員

警訊說明:(回收/矯正原因描述)

原廠發現特定型號/批號的新格導引線(Synchro Guidewires)發生聚四氟乙烯(Polytetrafluoroethylene, PTFE)塗層損壞的頻率增加,分析結果顯示可能是選配的導引器配件(Synchro Introducer Accessory)使用回載方式(Backloading)而造成塗層損壞問題,無需退回任何產品。

導引器配件供應商的製造流程存在差異,使得導引器配件比原預期有更銳利的邊緣,當醫生使用「回載方式」時,這些邊緣可能會導致PTFE塗層剝落,若剝落之塗層進入患者血管系統有可能產生潛在危害,包括術後組織壞死、短暫性腦缺血發作 (transient ischemic attack, TIA)、中風或死亡,PTFE栓塞也可能誘發慢性發炎。

為避免上述問題發生,請使用者不要使用回載方式(如Figure 1),應使用正確的加載方式(Figure 2)。迄今為止,尚未接獲患者受到傷害的案件。原廠已改進導引器配件,以降低導引器配件對導引線上PTFE塗層造成損壞的可能性。


Figure 1、「回載方式」指的是將導引線從導引器的遠端口進入,而不是從近端口進入的做法。


Figure 2、使用導引器的正確導引線加載方式

國內矯正措施:

經查,國內進口之受影響型號/批號數量共1,710個,美商史賽克(遠東)有限公司台灣分公司於113430日通知經銷商,請所有客戶立即檢查庫存是否有受影響產品,並提供安全通知給受影響之客戶;受影響產品仍可使用,惟務必依據安全警訊通知的指示操作(參閱附件)。前述矯正措施預計於113930前日完成。