跳到主要內容區塊

主題專區

::: 目前位置:首頁

託鏗血管攝影導管回收警訊(國內無進口受影響產品)
| 發布日期:2025-07-01 | 更新日期:2025-07-01 發布單位:醫療器材及化粧品組

託鏗血管攝影導管

回收警訊

(國內無進口受影響產品)

許可證字號:衛署醫器輸字第009437

產品英文名稱:TORCON NB ANGIOGRAPHIC CATHETER

受影響規格/型號/批號,及其UDI-DI

型號

批號

UDI-DI

HNBR5.0-35-100-P-NS-C2

(國內未進口受影響產品)

10827002087962

HNBR5.0-35-100-P-NS-DAV

10827002086996

HNBR5.0-35-100-P-NS-H1

10827002084213

HNBR5.0-35-100-P-NS-JB2

10827002088099

HNBR5.0-35-100-P-NS-SIM1

10827002084718

HNBR5.0-35-100-P-NS-SIM2

10827002084220

HNBR5.0-35-100-P-NS-SIM3

10827002085975

HNBR5.0-35-100-P-NS-VERT

10827002114835

HNBR5.0-35-125-P-NS-C2

10827002128993

HNBR5.0-35-125-P-NS-VANSCHIE1

10827002357331

HNBR5.0-35-125-P-NS-VANSCHIE4

10827002357362

HNBR5.0-35-40-P-NS-KMP

10827002097510

HNBR5.0-35-40-P-NS-VANSCHIE3

10827002314082

HNBR5.0-35-65-P-NS-C1

10827002084411

HNBR5.0-35-65-P-NS-C2

10827002084428

HNBR5.0-35-65-P-NS-C3

10827002084435

HNBR5.0-35-65-P-NS-KMP

10827002114583

HNBR5.0-35-65-P-NS-RC1

10827002084695

HNBR5.0-35-65-P-NS-RC2

10827002084701

HNBR5.0-35-65-P-NS-RDC

10827002085968

HNBR5.0-35-65-P-NS-RIM

10827002084442

HNBR5.0-35-65-P-NS-VANSCHIE1

10827002137957

HNBR5.0-35-65-P-NS-VANSCHIE2

10827002137940

HNBR5.0-35-65-P-NS-VANSCHIE3

10827002137926

HNBR5.0-35-65-P-NS-VANSCHIE4

10827002137834

HNBR5.0-35-65-P-NS-VANSCHIE5

10827002137933

HNBR5.0-35-80-P-NS-C2

10827002112022

HNBR5.0-38-100-P-NS-DAV

10827002085814

HNBR5.0-38-100-P-NS-JB1

10827002085838

HNBR5.0-38-100-P-NS-KMP

10827002114576

HNBR5.0-38-100-P-NS-SHETTY-050181

10827002114750

HNBR5.0-38-100-P-NS-SIM1

10827002085845

HNBR5.0-38-100-P-NS-SIM2

10827002084206

HNBR5.0-38-100-P-NS-VERT

10827002114842

HNBR5.0-38-125-P-NS-H1

10827002113784

HNBR5.0-38-125-P-NS-VERT

10827002121703

HNBR5.0-38-125-P-NS-VTK

10827002116426

HNBR5.0-38-40-P-NS-KMP

10827002095486

HNBR5.0-38-65-P-NS-C2

10827002084473

HNBR5.0-38-65-P-NS-H1

10827002113951

HNBR5.0-38-65-P-NS-KMP

10827002094694

HNBR5.0-38-65-P-NS-MPA

10827002111957

HNBR5.0-38-65-P-NS-RDC

10827002085784

HNBR5.0-38-65-P-NS-RIM

10827002100821

HNBR5.0-38-65-P-NS-SHK1.0

10827002114767

HNBR5.0-38-65-P-NS-SIM1

10827002086774

HNBR5.0-38-65-P-NS-SIM2

10827002086255

HNBR5.0-38-65-P-NS-SIM3

10827002088495

HNBR5.0-38-65-P-NS-VS1

10827002114866

HNBR5.0-38-80-P-NS-C1

10827002112039

HNBR5.0-38-80-P-NS-CHG2.5

10827002114309

HNBR5.0-38-80-P-NS-CHG3

10827002135731

HNBR5.0-38-80-P-NS-MIK

10827002093123

HNBR5.0-38-80-P-NS-MPB

10827002114699

HNBR5.0-38-80-P-NS-VS1

10827002111896

HNBR5.0-38-80-P-NS-VS2

10827002111902


發布對象:醫療專業人員

警訊說明:(回收/矯正原因描述)

原廠透過四起客戶投訴發現,特定批號產品中的導管出現尖端分離。進一步調查顯示,該問題是由於某位特定操作員的製程偏差造成的;因而回收含有該操作員生產零件的相關產品。

國內矯正措施:

經查,台灣曲克股份有限公司未進口警訊所述受影響產品。

廠商聯繫資訊:

公司名稱台灣曲克股份有限公司

聯絡電話02-66283542

聯絡人電子郵件stacey.chang@cookmedical.com

相關警訊來源(網址)

美國FDAhttps://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfres/res.cfm?id=213776